4-53
After system activation, the
activated functions are not
consistent with the users
configurations.
The factory activation code is
wrong. Or, the existing
configurations and the
configurations to be added
which are provided for the
factory are not complete.
Check the existing configurations
and the configurations to be added
again. Request the factory to
generate activation code again.
4.4.3 Change the Software Configuration of Electronic
Flowmeters
The software configuration of electronic flowmeters needs to be reset by the service engineer
when:
The user wants to change the existing electronic flowmeters.
The user wants to change the existing software configuration of electronic flowmeters.
The existing software configuration of electronic flowmeters does not meet the user’s
need.
NOTE
z Before changing the software configuration of electronic flowmeters, check the
number of tubes the user requires and the flowmeter standard complied (tube
order).
4.4.3.1 Check the Software Configuration of Electronic Flowmeters
Before resetting the software configuration of electronic flowmeters, check the configuration
information of the electronic flowmeters to be changed. The following table lists the
configuration information of electronic flowmeters including tube order and color which a
specific standard matches with.