6-33
NOTE
z Remove the vaporizer before tipping the anesthesia machine. If a vaporizer is
inverted, it must be set to 5% and purged for 30 minutes with a 5 L/min flow.
z Before tipping the anesthesia machine, maneuver it to coarse floor. Then step
down the two brakes to fix the machine. Do not over tip the anesthesia machine to
avoid personal injury or machine damage.
6.2.30 Replace the Drawer
Fully pull out the drawer. Hold the drawer sides by both hands. Depress the buckles inward
on both sides of the drawer to remove the drawer. Unscrew the three screws on the rail to
remove the rail. Remove the other rail similarly.
The following picture shows the appearance of drawer.
Buckle