3.12 Electrical Safety Tests
1. Perform leakage current test by using certified (such as UL, CSA or AMAI) test devices.
Make sure that the test result is not greater than 500 µA.
2. Make sure that the impedance between the protective grounding terminal of the power
cord and any exposed metal enclosure is less than 0.2Ω.