4-10 
3 
Breathing 
system leak 
test in manual 
ventilation 
mode  
Check the pneumatic circuit in manual ventilation mode 
for leaks, including APL valve, check valve, sodalime 
canister, patient tubes, flow sensors and their connectors. 
After each 
service or at 
the time of 
return visit 
4 
Check the 
sensors’ zero 
points 
Check if the zero points of all the flow sensors and 
pressure sensors inside the machine are within the normal 
range so as to determine when to replace the monitor 
board. 
After each 
service or at 
the time of 
return visit 
5 
Check the flow 
sensor 
accuracy 
1. Check if the measurements made by the flow sensors 
inside the machine are the same. 
2. Check if the measurement made by any flow sensor 
inside the machine is accurate.   
3. Check the effectiveness of flow calibration (factory) 
result. 
After each 
service or at 
the time of 
return visit 
6 
Check the 
pressure sensor 
accuracy 
1. Check if the measurements made by the pressure 
sensors inside the machine are the same. 
2. Check if the measurement made by any pressure sensor 
inside the machine is accurate. 
3. Check the effectiveness of pressure calibration (factory) 
result. 
After each 
service or at 
the time of 
return visit 
7 
Check the 
electronic 
flowmeter 
accuracy 
1. Check if the measurement made by the electronic 
flowmeter is normal. 
2. Check the effectiveness of electronic flowmeter 
calibration result. 
After each 
service or at 
the time of 
return visit 
 
 
4.2.1 Check the Mechanical Ventilation Mode 
 
NOTE 
z  The main function of the anesthesia machine is to provide breathing 
support—mechanical ventilation which complies with the doctor’s settings to the 
patient. The tests in this section are performed aiming to ensure that the machine is 
able to provide normal mechanical ventilation. 
z  The tests can help to judge if the machine operates normally.