Mindray WATO EX-65 - CO2 Calibration (Factory)

Mindray WATO EX-65
322 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-37
O2 supply pressure is insufficient
(lower than 200 kPa).
Change or connect the gas supply
and make sure that O2 supply
pressure is sufficient.
21% O2 calibration is not completed
before 100% O2 calibration.
Perform 21% O2 calibration
followed by 100% O2 calibration.
Calibration failure is
prompted about 3
minutes after
calibration is
started.
O2% count value is not within the
normal range (450 to 2700). Select the
[Maintenance] shortcut key
[Factory Maintenance >>]
[Diagnostic Test >>] [Display A/D
Channels] [Ventilator >>] to
access the [Display A/D Channels
Ventilator] menu.
Replace the O2 sensor.
4.3.8 CO2 Calibration (factory)
4.3.8.1 Preparations
Prepare the following before doing the calibration:
Gas cylinder: one or more cylinders filled with 3% , 4%, 5%, 6% , or 7% CO
2
T-shape connector
Samping line
4.3.8.2 Calibration Procedures
NOTE
z During the calibration, selecting [Calibrate] again does not take effect or exit the
calibration menu. Other operations than menu options are disabled until the end of
calibration.
Calibrate as follows:
1. Make sure that the CO2 module is already warmed up.
2. Select the [Maintenance] shortcut key [Factory Maintenance >>] enter the
required password [Module Cal. >>] [Gas Module Cal. >>] [CO2 Module
Cal.].
3. Check the airway and make sure that there are no occlusions or leaks.
Vent the sampling line to the air and check that the current rate is approximately
150 mL/min.If the deviation is great, it means that the airway is occluded. Check
the airway for occlusions.