4-56
4.4.4 Load Gas Module or BIS Module
The anesthesia machine provides online upgrade of gas modules and BIS module.
4.4.4.1 Check the Corresponding Relationship between the Loaded
Module and its Software Function
The following table lists the modules which can be loaded and their respective functions.
Module name Functional description Loading name of module
software function
AG module Provides AG and CO2 concentration
monitoring and O2 concentration monitoring
as well if O2 module is configured.
AG
M02B CO2
module
Provides CO2 concentration monitoring CO2
Mainstream CO2
module
Provides CO2 concentration monitoring CO2
Microstream
CO2 module
Provides CO2 concentration monitoring CO2
BIS module Provides BIS monitoring BIS
4.4.4.2 How to Load the Software Function of Gas Modules or BIS Module
To load the software function of the gas modules or BIS module:
1. Select the [Maintenance] shortcut key [Factory Maintenance >>] enter the
required password [Select Module >>] to open the module selection menu. Select
one or more from [CO2], [AG] and [BIS].
2. After the selection, restart the anesthesia machine or re-plug in the module to activate
the added module.