5-24
8. Turn on the N2O pipeline supply and record the reading on the test pressure gauge. If
the difference between this reading and the reading on the N2O pipeline pressure
gauge is more than 0.1 MPa (1bar), it indicates that the N2O pipeline pressure gauge is
damaged. Handle this problem as described in the troubleshooting table.
9. Insert the pulled-out tube properly.
III AIR supply inlet assembly:
1. Turn off the pipeline gas supply. Pull out No.50 PU tube which connects the AIR
supply inlet assembly to other assembly. The end of the tube which connects the AIR
supply inlet assembly is not pulled out but the other end is.
2. Connect the pulled-out tube end to the test pressure gauge through 3106-04-06 adapter
connector, as shown below.
3. Turn on AIR pipeline supply. If the reading on the test pressure gauge is not within the
range of 0.2±0.05 MPa (namely 2±0. 5bar), adjust the regulator to cause the reading on
the test pressure gauge to reach 0.2 MPa (namely, 2bar). For operations of the
regulator, refer to section 5.3.2.4Adjust the Regulator of the Pipeline Gas Supply
Inlet Assembly. Record the reading on the AIR pipeline pressure gauge.