5-30 
The flowmeter float 
indicates inaccurate value 
or remains unmoved.   
The float flowmeter is 
damaged. 
Replace the float flowmeter. 
The knob of the flow 
regulator gets loose. 
The flow regulator is 
damaged. 
Replace the flow regulator. 
The O2-NO link system is 
ineffective. 
The O2-N2O chain linkage of 
the flow regulator is damaged.
Replace the flow regulator. 
N2O supply cannot be cut 
off in case of O2 supply 
failure. 
The O2-N2O cut-off valve is 
damaged. 
Replace the O2-N2O cut-off valve 
assembly. 
 
5.3.3.1 Leak Test of the O2 Flush Button Assembly 
Perform a leak test of the O2 flush button assembly by using the following tools:   
  Negative pressure ball (quantity:1) 
  3106-06-00 adapter connector (quantity:1) 
  PU tube (6X100) (quantity:1) 
Test procedures: 
1.  Turn off the pipeline gas supplies and bleed the residual pressure through O2 flushing.   
2.  Pull out No.52 PU tube which connects the O2 flush button assembly to the ACGO 
assembly. The end of the tube which connects the O2 flush button assembly is not 
pulled out but the other end is. 
3.  Flatten the negative pressure ball to remove the gas inside. Then re-install the gas 
outlet switch of the negative pressure ball properly. Connect the other end of the 
negative pressure ball to the pulled-out tube through 3106-06-00 adapter connector. 
4.  Release the negative pressure ball as shown below. If the negative pressure ball is fully 
expanded within 30 s, it indicates that the O2 flush button assembly is damaged. 
Handle this problem as described in the troubleshooting table.