5-68 
The cable between the 
monitor board and the 
main control board gets 
loose. 
Properly insert the cable between the 
monitor board and the main control 
board. 
Exiting Standby fails. 
The monitor board 
hardware selftest is failed.
Return the monitor board to factory 
for repair. 
The buttons are 
ineffective. 
Replace the keyboard. The buttons malfunction. 
The cable gets loose.  Properly insert the cable between the 
keyboard and button board. 
The isolation transformer 
buzzes. 
The anesthesia machine is 
powered by 220V AC. 
Resort to 110 V power supply 
Alarm messages are 
displayed on the screen but 
without alarm sound. 
 
The button board or 
speaker is damaged. 
Return the button board to factory 
for repair or replace the speaker. 
The control knob is 
ineffective. 
Replace the control knob. Operating the control knob is 
not responded. 
 
The button board is 
damaged. 
Return the button board to factory 
for repair. 
The cables connected to 
the network connection 
board get loose. 
Properly insert the cables. 
The network cable is too 
long. 
Shorten the network cable. 
Recommended cable length is 
approximately 1.5 m. 
Network connection is 
failed. 
The network cable is used 
incorrectly. 
The network cable has two linear 
orderings which should be 
differentiated. 
The bag/mechanical 
ventilation switch is 
faulty or the ACGO 
switch is turned on. 
Check the screen to see if the 
anesthesia machine is in mechanical 
ventilation mode and if there is an 
alarm triggered. 
No gas is outputted through 
the valve in mechanical 
ventilation mode. 
The valve cannot be 
opened.  
1. Set tidal volume to maximum.   
2. Switch between standby and 
mechanical statuses or between 
manual and mechanical statuses 
repeatedly.  
3. Replace the pneumatic circuit 
block.