13
1 Before Using This Product
No. 99MAF600B
5
Display [5] in ones place using the same procedure described in steps
3
and
4
.
6
Slowly bring the ball of the probe into contact with the 25 mm gauge block until a beep sound is made.
» Reference setup is complete.
Tips
For measurements using a dial test indicator or dial indicator, bring the stylus into contact with the gauge block,
and then press to nish the setup.
INC reference setup
Tips
If the power to the product is cut off by removing the batteries or the AC adapter (optional accessory) cable, the
reference setting will be erased.
1
Press .
2
Slowly bring the ball of the probe into contact with the desired reference until a beep sound is made.
» Reference setup is complete.