EasyManua.ls Logo

Molecular Devices FlexStation 3 - Fluorescence Intensity Performance; Fluorescence Polarization Performance

Molecular Devices FlexStation 3
173 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Specifications
0112-0127 B 157
Table B-4: Fluorescence Intensity Performance
Item Specification
Sensitivity Top Read
< 5 pM FITC, 1 fmol/200 μL (96)
< 20 pM, 2 fmol/100 μL (384)
Wavelength range 250 nm to 850 nm
Wavelength selection Monochromators, tunable in 1 nm increments
Bandwidth (excitation,
emission)
9 nm, 15 nm
Number of
excitation/emission pairs
per plate
4
Dynamic range 106 in 96-well black plates: auto gain circuitry
System validation Self-calibrating with built-in fluorescence calibrators
Light source Xenon flash lamp (1 joule/flash)
Average lamp lifetime 2 years normal operation
Detector Photomultiplier
Table B-5: Fluorescence Polarization Performance
Item Specification
Wavelength range 400 nm to 750 nm
Wavelength selection Monochromators, tunable in 1 nm increments
Bandwidth (excitation,
emission)
9 nm, 15 nm
Precision < 5 mP standard deviation at 1 nM fluorescein
in 96-well and 384-well microplates
FlexStation_manual.book Page 157 Tuesday, June 25, 2013 10:22 AM