Innity
®
Enteral Feeding Pump Operator’s Manual
18
2. DIRECTIONS FOR USE
The Battery
Charging the Battery
To charge the battery, insert plug from AC
Adapter/Charger into port on the left side
of pump. Plug charger into a wall outlet.
Pump will charge whether it is
turned o or running.
When charger is plugged in and charging, the
plug symbol will appear in the display and the
fuel gauge will display four bars scrolling from
E to F. This pattern continuously repeats while
pump is charging (Figures 2-28a through 2-28e).
When bars stop scrolling pump is done charging.
To check the status of battery during
charging, disconnect charger from
pump and turn on pump.
NOTE: If pump shuts down due to low
battery, charging less than 10 minutes may
result in an ER99 alarm. Fully charge battery
if depleted (see page 22 to clear alarm).
a
b
c
d
e
Figure 2-28 Fuel Gauge

Table of Contents

Other manuals for Moog Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Moog Infinity and is the answer not in the manual?

Moog Infinity Specifications

General IconGeneral
ModelInfinity
TypeInfusion Pump
Occlusion Pressure SettingsLow, Medium, High
AlarmsOcclusion, Low battery
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C
ManufacturerMoog
Infusion ModesContinuous, Intermittent
DisplayLCD
CompatibilityStandard infusion sets