Enteral Feeding Pump Operator’s Manual
iv
LIST OF WARNINGS AND CAUTIONS
Be sure to carefully read the following warnings and cautions. Failure to
review all the warning and cautions can cause harm to the operator or
patient. Healthcare personnel who provide training to lay users should be
sure to review all the warning and cautions with the lay user. If there is a
change in the performance of the pump, contact customer service.
WARNINGS
4 Use ONLY feeding solutions prescribed by the responsible
physician, registered dietitian, registered nurse, or other licensed
practitioner. Failure to do so may result in harm to the patient.
4 Use ONLY Innity brand disposable sets to ensure proper uid
delivery. Others will not deliver the correct dose, may allow dangerous
free-ow conditions, and may generate hazardous uid pressures
which may activate occlusion alarms at unpredictable pressures.
4 Low ow rates, combined with high dose settings may exceed the life
of the disposable set. The disposable bag set must be replaced every
24 hours and the spike sets must be replaced every 48 hours (max1.5 L)
to maintain delivery accuracy, allow proper air and occlusion sensing,
and prevent growth of bacteria.DO NOT PROGRAM RATE AND DOSE
COMBINATIONS WHICH EXCEED A 24 HOUR FEED REGIMEN FOR THE
DISPOSABLE BAG SETS OR A 48 HOUR FEED REGIMEN FOR THE SPIKE SETS.
4 The battery capacity is an approximation. If you are unsure that
enough capacity remains for your intended use, recharge it. Failure
to do so may result in delayed or under-delivered therapy.
4 To avoid electrical shock, never clean pump with charger
plugged into an outlet or when pump is on.
4 To avoid electrical shock, make sure the Innity AC Adapter/Charger
is completely dry before plugging into an electrical outlet.
4 Do not use Innity Enteral Feeding Pump for delivery of
non-enteral solutions. Serious injury may result.
4 Proper operation of pump requires door to be closed and latched.
Make sure door is closed and latched when motor is running.
4 If an error occurs (ER01 - ER99 will appear in display), check all settings
and volume counters and verify before starting a feeding. Failure
to do so may result in delayed or under-delivered therapy.

Table of Contents

Other manuals for Moog Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Moog Infinity and is the answer not in the manual?

Moog Infinity Specifications

General IconGeneral
ModelInfinity
TypeInfusion Pump
Occlusion Pressure SettingsLow, Medium, High
AlarmsOcclusion, Low battery
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C
ManufacturerMoog
Infusion ModesContinuous, Intermittent
DisplayLCD
CompatibilityStandard infusion sets

Related product manuals