© Copyright 2020 Moog, Inc.
31
5. CLEANING
5. CLEANING
WARNING: To avoid electrical shock, never clean pump or Innity AC Adapter/
Charger with charger plugged into an outlet or pump turned on.
WARNING: Make sure the Innity AC Adapter/Charger is
completely dry before plugging into an electrical outlet.
To clean the Innity enteral feeding pump:
Pump may be cleaned with warm soapy water (standard dish soap) and a
nonabrasive sponge or soft cloth. Use a cotton swab to clean pathways of cassette
receptacle and to remove teal silicon residue from pump wheel roller pins.
OR - Apply one of the following solutions for approximately 10
minutes, then wipe pump clean with a damp cloth or sponge:
4 5% bleach and water solution
4 Multipurpose household disinfectant cleaner
Rinse pump by holding under a stream of
warm water. Then, dry with a clean cloth.
NOTE: Clean pump as needed. If the pump is used on
more than one patient, clean between each patient.
NOTE: Avoid harsh cleaners/disinfectants. They
may cause damage to pump surfaces and may
aect pump functionality. Moog has tested
the following disinfectants for acceptability for
cleaning/disinfecting the Innity pump. These
cleaning products are categorized below:
Acceptable Unacceptable
Metrex CaviCide Ball Ruthless
5% Bleach and Water Solution Wex-Cide 128
Warm Soapy Water (Standard Dish Soap) Alcohol
Vesphene Ilse
To clean the Carrying Packs:
The Innity packs are machine-washable. Use cold water
and gentle cycle, whenever possible. Hang to dry.
To clean the AC Adapter/Charger:
Innity AC Adapter/Charger normally does not require cleaning. When
desired, a dry or slightly damp cloth may be used to clean the outside surface
of AC Adapter/Charger while it is disconnected from the wall outlet.
Figure 5-1 Rinsing Pump
Under Stream of Water

Table of Contents

Other manuals for Moog Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Moog Infinity and is the answer not in the manual?

Moog Infinity Specifications

General IconGeneral
ModelInfinity
TypeInfusion Pump
Occlusion Pressure SettingsLow, Medium, High
AlarmsOcclusion, Low battery
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C
ManufacturerMoog
Infusion ModesContinuous, Intermittent
DisplayLCD
CompatibilityStandard infusion sets

Related product manuals