EasyManuals Logo

Moog Infinity User Manual

Moog Infinity
113 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
Innity
®
Enteral Feeding Pump Operator’s Manual
10
2. DIRECTIONS FOR USE
Operating the Pump
A Single Feeding Example
CAUTION: Please refer to pages v-vi for additional usage considerations.
If you want to deliver 500 ml of enteral feeding solution at
a rate of 120 ml/hr and then stop the pump:
Key Sequence for “A Single Feeding Example
120 ml/h 500 ml
Step 1:
Prepare the disposable set with enteral feeding solution as described in
the Priming and Loading the Disposable Set instructions (page 7).
Step 2:
Hang feeding bag or container so that the bottom of bag is at
or above the level of the pump door (refer to page 37).
OR - If an Innity carry pack is to be used, load pump and
feeding bag into the proper compartments, securing pump,
bag, and tubing with the pack’s straps (Pages 33-34).
Step 3:
Turn pump on by pressing the ON/OFF
key. After running the self test, display
will show last programmed rate.
Step 4:
Press the + or - key to change the rate to 120 ml/hr.
Hold down either key to change rapidly (Figure 2-11).
Step 5:
Press the RATE/DOSE key to display dose. Press the +
or - key until a dose of 500 ml is displayed (Figure 2-12).
Figure 2-11 Program
Rate to 120 ml/hr
Figure 2-12 Program
Dose to 500 ml

Table of Contents

Other manuals for Moog Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Moog Infinity and is the answer not in the manual?

Moog Infinity Specifications

General IconGeneral
ModelInfinity
TypeInfusion Pump
Occlusion Pressure SettingsLow, Medium, High
AlarmsOcclusion, Low battery
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C
ManufacturerMoog
Infusion ModesContinuous, Intermittent
DisplayLCD
CompatibilityStandard infusion sets

Related product manuals