Enteral Feeding Pump Operator’s Manual
10
2. DIRECTIONS FOR USE
Operating the Pump
A Single Feeding Example
CAUTION: Please refer to pages v-vi for additional usage considerations.
If you want to deliver 500 ml of enteral feeding solution at
a rate of 120 ml/hr and then stop the pump:
Key Sequence for “A Single Feeding Example
120 ml/h 500 ml
Step 1:
Prepare the disposable set with enteral feeding solution as described in
the Priming and Loading the Disposable Set instructions (page 7).
Step 2:
Hang feeding bag or container so that the bottom of bag is at
or above the level of the pump door (refer to page 37).
OR - If an Innity carry pack is to be used, load pump and
feeding bag into the proper compartments, securing pump,
bag, and tubing with the pack’s straps (Pages 33-34).
Step 3:
Turn pump on by pressing the ON/OFF
key. After running the self test, display
will show last programmed rate.
Step 4:
Press the + or - key to change the rate to 120 ml/hr.
Hold down either key to change rapidly (Figure 2-11).
Step 5:
Press the RATE/DOSE key to display dose. Press the +
or - key until a dose of 500 ml is displayed (Figure 2-12).
Figure 2-11 Program
Rate to 120 ml/hr
Figure 2-12 Program
Dose to 500 ml