Innity
®
Enteral Feeding Pump Operator’s Manual
22
Alarms
ER01 - ER99 (Figure 3-1)
Pump will sound a continuous beep and display
will read ER followed by a two digit number or ERR
followed by a letter to indicate a self-test has failed.
To clear alarm: Turn the pump
o and then back on.
WARNING: If an error occurs
(ER01 - ER99 will appear in display), all settings and volume counters
should be checked and veried before starting a feeding.
LOAD SET (Figure 3-2)
Pump will sound a dual tone beep repeatedly
and display will read LOAD SET.
Why: This alarm will sound when pump attempted
to run with the set improperly loaded or missing.
To correct the problem: Press the RUN/
PAUSE key to silence alarm and place pump
in pause mode, then check the following:
1. An Innity disposable set is loaded into
pump properly and door is closed.
2. Check cassette receptacle for cleanliness,
especially around pressure sensors.
If cleaning is necessary, refer to
CHAPTER 5 - CLEANING (Page 31).
LOW BATT (Figure 3-3)
When 1 hour or less of battery life is available,
display will ash LOW BATT every 3 seconds and
pump will beep every 2 seconds to indicate battery
is low and pump will soon stop running.
To correct the problem: Plug in AC
Adapter/Charger to recharge battery.
3. ALARMS, MESSAGES, AND INDICATIONS
Figure 3-1 ER01 - ER99 Alar ms
Figure 3-2 LOAD SET Alarm
Figure 3-3 LOW BATT Alarm