Innity
®
Enteral Feeding Pump Operator’s Manual
28
4. TROUBLESHOOTING GUIDE
4. TROUBLESHOOTING GUIDE
NO FOOD Alarm
If re-priming of the pump delivery set is required, it is advisable to disconnect
the enteral adapter from the patients feeding tube while re-priming.
ER01 - ER99 Alarms
Is bag of pump
delivery set empty?
Rell bag, re-prime delivery set and resume feeding.
NO
Is air present in
the tubing?
Has all air been removed from bag of pump delivery set?
IF NO: Manually prime tubing by
turning bag upside down, gently
squeezing lower half of bag while
pinching the teal colored tubing
just below the
drop symbol.
Hold this position until all air is
removed from bag and tubing.
IF YES: Press and
hold the PRIME key
to move the air in the
tubing past pump
delivery set cassette
and resume feeding.
Let formula sit for 10-15 minutes before preparing pump delivery
set for feeding. If foam is present in tubing, ush by priming pump
delivery set either manually or with the pump and resume feeding.
If alarm continues, contact Moog Customer
Service at 800.970.2337 for assistance.
YES
YES
Is formula blenderized
or aggressively mixed?
YES
NO
NO
Are air bubbles
trapped inside
the cassette?
Remove tubing from pump and manually prime tubing by turning
bag upside down, gently squeezing lower half of bag while pinching
the teal colored tubing just below the drop symbol. Hold this position
until all air has passed through the teal segment of the tubing.
YES
NO
If alarm continues, contact Moog Customer
Service at 800.970.2337 for assistance.
Was pump door
open while pump
was turned on or
while running?
Turn pump o. Unload and reload delivery
set cassette. Make sure door is closed
securely, then turn pump back on.
YES / UNSURE
NO

Table of Contents

Other manuals for Moog Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Moog Infinity and is the answer not in the manual?

Moog Infinity Specifications

General IconGeneral
ModelInfinity
TypeInfusion Pump
Occlusion Pressure SettingsLow, Medium, High
AlarmsOcclusion, Low battery
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C
ManufacturerMoog
Infusion ModesContinuous, Intermittent
DisplayLCD
CompatibilityStandard infusion sets