© Copyright 2020 Moog, Inc.
13
2. DIRECTIONS FOR USE
Next Feeding:
Turn pump on. Previous settings for rate, dose, and feed interval
will be saved in memory. If no changes are required, verify
each setting is correct and restart pump running.
NOTE: Rate, dose, and feed interval settings can be locked so
that changes cannot be made. See the User Preference Settings
section (Page 19) for instructions to change this feature.
Operating the Pump
An Interval Feeding Example
CAUTION: Please refer to pages v-vi for additional usage considerations.
If you want to deliver 100 ml of enteral feeding solution at a
rate of 50 ml/hr, and repeat this feeding every 6 hours:
Key Sequence for “An Interval Feeding Example”
50 ml/h 100 ml 6.00- hr
Step 1:
Prepare the disposable set with enteral feeding solution as described in
the Priming and Loading the Disposable Set instructions (Page 7).
Step 2:
Hang feeding bag or container so that the bottom of bag is at
or above the level of the pump door (refer to page 37).
OR - If an Innity pack is to be used, load the pump and feeding bag into the proper
compartments, securing pump, bag and tubing with the pack’s straps (see Page 32).
Step 3:
Turn pump on by pressing the ON/OFF key. After running the
self test, display will show last programmed rate.