Enteral Feeding Pump Operator’s Manual
12
2. DIRECTIONS FOR USE
To restart pump when in paused mode, press the RUN/PAUSE key. The
pump will restart the feeding at the point where it was paused.
NOTE: ANY CHANGES TO PUMP SETTINGS DURING A FEEDING CYCLE WILL CAUSE
PUMP TO START A NEW FEEDING; IT WILL NOT START WHERE IT LEFT OFF.
Feeding Completion:
When the dose has been completely delivered, pump will stop running, and display
will read DOSE DONE. Pump will be in a pause mode until it is turned o or feeding
is restarted. To clear DOSE DONE, press and hold the ON/OFF key for 1.5 seconds to
turn pump o, or press the RUN/PAUSE key to put pump in pause mode. Adjustments
to settings can be made at this time, or press the RUN/PAUSE key a second time
to start another feeding using same rate, dose, and feed interval settings.
If you would prefer the alarm to sound when the
dose is complete, you can change the pump dose
complete alarm setting to BEEP WHEN DONE. See
the User Preference Settings section (Page 19) for
instructions to change this setting. When set to BEEP
WHEN DONE, the pump will stop at the end of the
feeding, beep intermittently, and display will read
DOSE DONE (Figure 2-18). To silence alarm when
the dose is complete, press and hold the ON/OFF
key for 1.5 seconds to turn pump o, or press the
RUN/PAUSE key to put pump in pause mode.
NOTE: To program an innite dose, press
and hold the + key until display reads INF.
When an innite dose has been programmed, pump will not alarm DOSE
DONE. Pump will run continuously as long as feeding solution remains in
the bag. When empty, pump will alarm NO FOOD or NO FLOW IN.
Figure 2-18 DOSE DONE

Table of Contents

Other manuals for Moog Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Moog Infinity and is the answer not in the manual?

Moog Infinity Specifications

General IconGeneral
ModelInfinity
TypeInfusion Pump
Occlusion Pressure SettingsLow, Medium, High
AlarmsOcclusion, Low battery
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C
ManufacturerMoog
Infusion ModesContinuous, Intermittent
DisplayLCD
CompatibilityStandard infusion sets

Related product manuals