11
2. DIRECTIONS FOR USE
NOTE: For a single feeding the feed interval
setting should be programmed to NONE (Figure
2-13). When you have nished setting rate and
dose check the display to see if it reads FEED
INTERVAL. If it does, press the FEED INT key, then
press the CLEAR key. Display will read NONE.
Step 6
Connect end of disposable set to the patient’s
enteral feeding tube (Figure 2-14).
Step 7
Press the RUN/PAUSE key. Pump will begin running.
Display will show programmed rate and the arcs
around the run symbol will rotate (Figure 2-15).
While pump is running, the following may
be viewed by pressing the appropriate
key: Rate, Dose, and Feed Interval.
To view the amount delivered in the current
feeding, press the VOL/TOTAL key (Figure 2-16).
This counter will reset itself when the previously
programmed dose has been completed and a new
feeding is started, or if a feeding is interrupted
and Rate, Dose or Feed Interval is changed.
To view the amount delivered over the course
of several feedings press the VOL/TOTAL key
a second time. Display will read TOTL then the
amount. This counter never resets itself. It can
only be reset by the user (Figure 2-17).
While pump is running, the settings cannot be
changed and the Prime feature is disabled.
If you want to stop pump at any time, press the RUN/
PAUSE key. Display will show rate, and pump will
be in pause mode. Pump will save the memory of
where it was in the feeding before being paused.
Or, turn pump o by pressing the ON/OFF key.
Figure 2-16 Amount
Currently Delivered
Figure 2-17 Total
Amount Delivered
Figure 2-13 Clear Feed Interval
Figure 2-15 Pump is Running
Figure 2-14 Connect disposable
set to feeding tube