8. IMPORTANT PEDIATRIC CONSIDERATIONS
Innity can be used on pediatric patients if the specications of pump meet delivery
requirements of the patient. The specications of primary importance are:
4 The ow rate range of Innity is 0.1 ml/hr to 10 ml/hr in 0.1 ml/hr
increments and 10 ml/hr to 600 ml/hr in 1 ml/hr increments.
4 The pump and administration set (“pump system”) is accurate
to ±5% per volume delivered over one hour for ow rates
of 5-600 mL/h, under the following conditions:
8 utilizing Moog Innity disposable sets
8 uid head height at +6.0 inches ± 0.3 inches (+15.24 cm ± 0.76 cm) at
95% condence and 90% reliability with respect to center of rotor.
Note: Head height eect on accuracy is shown below (Figure 8-1).
Figure 8-1 Head Height Eect on Accuracy
+2.5%
-2.5%
-12 in.
(-30.5 cm)
+12 in.
(+30.5 cm)
Normal
If these specications meet the required feed regimen, Innity can be
used to deliver enteral formulas to pediatric patients. ALWAYS VERIFY
RATE, DOSE AND FEED INTERVAL BEFORE PROCEEDING TO FEED.
8. IMPORTANT PEDIATRIC CONSIDERATIONS

Table of Contents

Other manuals for Moog Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Moog Infinity and is the answer not in the manual?

Moog Infinity Specifications

General IconGeneral
ModelInfinity
TypeInfusion Pump
Occlusion Pressure SettingsLow, Medium, High
AlarmsOcclusion, Low battery
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 60°C
ManufacturerMoog
Infusion ModesContinuous, Intermittent
DisplayLCD
CompatibilityStandard infusion sets

Related product manuals