Enteral Feeding Pump Operator’s Manual
20
2. DIRECTIONS FOR USE
To adjust these settings:
Key Sequence for “Adjusting User Preference Settings”
HIGH LOCK MUTE LITE ON
LOW UNLK BEEP LITE OFF
Step 1:
With pump turned o, press and hold for 1.5 seconds the + and the ON/
OFF keys at the same time. Pump will beep and then display the words
BEEP HIGH or BEEP LOW. Press the + key to change to HIGH or the - key to
change to LOW. Pump will give a triple beep when the + or - keys are pressed
and setting will change. If setting is as desired, do not press either key.
Step 2:
Press the PRIME key. Display will read UNLK (unlocked) or LOCK (locked).
Press the + key to change to LOCK (locked) or the - key to change to
UNLK (unlocked). If setting is as desired, do not press either key.
Step 3:
Press the PRIME key again. Display will read MUTE WHEN DONE or BEEP WHEN
DONE. Press the + key to change to BEEP WHEN DONE or the - key to change
to MUTE WHEN DONE. If setting is as desired, do not press either key.
Step 4:
Press the PRIME key again. Display will read LITE ON or LITE OFF.
Press the + key to change to LITE ON or the - key to change to
LITE OFF. If setting is as desired, do not press either key.
Step 5:
Press the ON/OFF to turn pump o. Setting changes will be saved automatically.
NOTE: Once user has changed the desired user preference setting,
it is not necessary to view the remaining user preference settings.
Each change will be automatically saved as it is made.
NOTE: Pump will turn o automatically after 2 minutes in the User
Preference Settings if the user does not turn it o before then.