Metallic and Plastic Parts Cleaning
Exterior Cleaning
Use a damp cloth to clean external covers and the patient cable. Do not use alcohol, solvents, or cleaning solutions.
These cleaning agents may damage the surfaces of the instrument.
Interior Visual Inspection
Check all connectors for loose, bent or corroded contact points
Check wire, harnesses and cables for signs of wear or deterioration.
Inspect leads for security of mounting, or deterioration.
Check terminals and connections for proper installation, loss or wear.
Check the identification nameplate and other decals for legibility.
Inspect chassis and covers for warping, bending, surface damage or missing captive hardware.
Check all screws and nuts for tightness or signs of stripped or crossed threads.
Check for any other form of mechanical damage, which may indicate a failure.
Preventative Maintenance Schedule:
Maintenance to be Performed
Perform every 3 mo. if unit is in heavy use.
CAUTION
Do not wipe over surfaces of nameplates or labels with abrasive cleaners or materials, as
this will eventually wear away the nameplate information. Do not use solvents to clean
plastic parts.