Section 2 Mode section
Section 2 of the log.txt file contains information regarding what happens prior to the actual data recording. This will
include entries for:
Mode description
Value entered in log.txt file
Main_Menu
0
Impedance
1
Waveform
2
ID
3
Recording
4
Configuration
5
DATETIME Menu
6
Language
7
Version
8
Set Date & Time
9
From the example above the Mode section contains the following information (please note that different recordings
will have different information):
04/05/2004 08:07:51 Mode 1
04/05/2004 08:08:08 Impedance 2
Impedance 3
Impedance 8
Impedance 1
Impedance 0
Impedance 1
Impedance 9
Impedance 11
Impedance 17
04/05/2004 08:08:23 Mode 3
04/05/2004 08:09:00 Mode 2
04/05/2004 08:09:13 Mode 2
04/05/2004 08:10:00 Mode 4
04/05/2004 10:02:49 Event
Mode 1 Impedance
Shows the impedance values, in K Ohms, of leads (RA, LA, LL, V1, V2, V3, V4, V5, V6) given during the
impedance test. The closer the value is to 0 the better the electrical connection to the patient.
Mode 2 Waveform
An entry of Mode 2 indicated that the Waveform display was accessed.
Mode 3 ID
An entry of Mode 3 (ID) means that the ID screen was accessed.
Mode 4 Recording
An entry of Mode 4 (Recording) means that the Record option was selected and that the monitoring session has
begun. The entry after this will be and Event indicating the start of the monitoring session.