Introduction
The H12+ Holter recorder generates several files when recording a monitoring session. One of these files, the log.txt
file, can be useful in determining problems or events that occurred during a monitoring session. This file is
accessible after the FLASH card data has been downloaded to the H-Scribe computer system. The log file will be
located in the following area on the H-Scribe computer and can be viewed using a text editor:
C:\USR\PATXX
Where XX is a number that is equivalent to the slot location of the stored Holter recording.
Below is a sample of the information in a log file:
08:07:32 Power up Serial Number:
04/05/2004 08:08:23 Mode 3
04/05/2004 08:09:00 Mode 2
04/05/2004 08:09:13 Mode 2
04/05/2004 08:10:00 Mode 4
04/05/2004 10:02:49 Event
04/05/2004 10:35:30 Lead Fault 638
04/05/2004 10:35:32 Lead Fault 0
04/05/2004 10:36:05 Lead Fault 126
04/05/2004 10:36:07 Lead Fault 0
04/05/2004 10:42:43 Lead Fault 023
04/05/2004 10:42:45 Lead Fault 0
04/05/2004 10:48:51 Lead Fault 023
04/05/2004 10:48:53 Lead Fault 0
The log file is separated into several sections, Power up section, Mode section and Fault section.
Section 1 Power up section
The first section of the log file is the power up section. This entry is a 1 line entry that contains the date, time, Power
up statement and the H12+ serial number. From the example above this line looks like:
04/05/2004 08:07:32 Power up Serial Number: 102401428160