10.8 Display of overdriven measured values
Narda FieldMan 107
10.8 Display of overdriven measured values
Overdriving can lead to damage or even destruction of a measuring probe. In addition,
during overdriving, the measurement results are undervalued due to the occurring com-
pression and thus recorded as too low. Measurement data stored in data records are
identified by override symbols.
A currently overdriven state is indicated by a red instead of a yellow display field in con-
nection with an upward pointing triangle symbol
Due to previous overdriving, the statistical measured values of MAX, MAXAVG and AVG
are also undervalued and are therefore displayed in red after overdriving has occurred.
In this case, the values in the measurement settings should be reset.