7.4 LEGAL NOTICES
Narda FieldMan 47
7.3.5 DEVICE
7.4 LEGAL NOTICES
In this submenu you will find information about the radio approvals.
⇒ For more information, see 13.5 Radio approval on page 119.
7.5 EXPERT SETTINGS
This submenu is probe-dependent and can only be selected if an HF probe or BFD-400
is connected. There are no adjustment options for other probes.
7.5.1 POST-PROCESSING
Parameter Default Description
Format Memory Card FORMAT Format SD card after confirmation
Automatic Shutdown 30 min Power off during inactivity
Calibration Reminder When due Can be activated on expiry or in ad-
vance, or generally deactivated
Factory Reset RESET Sets all adjustable parameters to the
as-delivered values
In this submenu, settings can be made for HF probes that require precise knowledge of
the prevailing field. If the frequency of a field source is known and is specified here as the
reference frequency, a more accurate display result can be obtained by applying the cor-
rection factor stored in the probe, or the percentage of the limit that has been reached
can be displayed in % relative to a safety standard (see 10.1 Post-Processing on
page 98).
HF probes are broadband probes and cannot measure selectively at the set frequency.
In the case of incorrect frequency specification or multifrequency fields, these settings
can therefore lead to incorrect results.