5.3 Explanation of Commands
DF1906
5-12
[:SOURce]:BURSt: SPACes <num space>
[:SOURce]:BURSt:SPACes?
Explanation: Sets/queries the stop wave number for burst oscillation.
Parameter: Stop wave number
0.5 to 256
Response: Returns the set stop wave number.
Nu meric data is transferred as ASCII byte in NR2 format.
ex) BURS:SPAC?
10.0
Remark: The stop wave number can be set in increments of 0.5 wave for burst oscillation
when both the mark wave number and space wave number are 128 or less.
[:SOURce]:PHASe <phase>
[:SOURce]:PHASe?
Explanation: Sets/queries the oscillation stopping phase for burst oscillation, etc.
Parameter: Phase [deg]
-180.0(-180deg) to +180.0(+180deg)
Response: Returns the set phase in [deg].
Nu meric data is transferred as ASCII byte in NR2 format.
ex) PHAS?
90.0
:OUTPut:SYNC:TYPE {PHASe|STATe}
:OUTPut:SYNC:TYPE?
Explanation: Sets/queries the SYNC OUT output type.
Parameter:
PHASe (Waveform synchronous output)
STATe (Oscillation state synchronous output)
Response: Returns PHAS or STAT.
ex) OUTP:SYNC:TYPE?
PHAS
:TRIGger[:SEQuence][:IMMediate] {0|1}
Explanation: Controls the trigger or gate signal.
Parameter:
In trigger mode
0 (No change)
1 (Trigger input)
In gate mode
0 (Gate off)
1 (Gate on)