2.2 Installation
5
WF1973/WF1974
c) Repacking
When repacking this equipment for transportation, etc., use a shipping carton of sufficient
strength and capacity to safely accommodate the equipment and hold its weight.
d) Options
The following options are available and can be purchased separately.
Multi-I/O cable (PA-001-1318)
This cable is used when using the multi-I/O connector on the rear panel.
A 2 meter multi-core shielded cable is connected to the mini-Dsub 15-pin connector. Since
the opposite side is cut off, process that side according to the connection destination.
For the connector's pin assignment and cable differentiation, ) p. 18.
Rack mount adapters
These adapters are for mounting the equipment on a 19-inch IEC, EIA standard rack, or
JIS standard rack.
Each type of adapter is available as a 1-unit and 2-unit model (for side-by-side mounting),
for a total lineup of 4 models.
2.2 Installation
a) Installation sites
Do not place the equipment with the rear panel facing down, because this may damage the
connectors and hinder ventilation.
Place the equipment on a flat surface such as a desk so that the four rubber feet and stands
on the bottom side rest on that surface.
b) Installation location conditions
This product uses a fan for forced-air cooling and features air intake and exhaust vents on the
side and rear panels for this purpose. To allow for ample air flow, be sure to maintain a gap of
at least 10 cm between the sides and rear of this product and walls or other obstructions.
Install this product in a location that meets the following conditions for temperature and
humidity ranges.
Operating conditions: 0 to 40ºC, 5 to 85%RH
Storage conditions: 10 to 50ºC, 5 to 95%RH
Further, a condensation-free environment must be ensured. For limitations related to
absolute humidity, refer to the specifications in this manual.
Do not install the WF1973/WF1974 in the following locations:
Location with flammable gas
An explosion may occur. Never install and use this product in such a location.
Outdoors, or location exposed to direct sunlight or near a fire or heat source
The full performance of this product may not be obtained, or failure may occur.
Location with corrosive gas, moisture, dust, or high humidity
This product may become corroded or fail.
Location near an electromagnetic field source, high-voltage device, or power line
This product may malfunction.
Location exposed to excessive vibration
This product may malfunction or fail.