EasyManua.ls Logo

NF WF1973 - Setting the Ramp Wave Symmetry

NF WF1973
169 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
64
4. BASIC OPERATION
MULTIFUNCTION GENERATOR
4.4.16 Setting the ramp wave symmetry
The waveform is assumed to be set to [Ramp] (ramp wave). For the waveform setting
method, ) p. 44.
The symmetry setting unit is % only. Setting and display with time is not possible.
a) Symmetry setting method
1. Select the [Symm] field and then press the
ENTER key to open the symmetry
input field.
Even if there are several setting screens,
the [Symm] field is always displayed on
the first page.
2. Select the digit to be changed with the left/right arrow keys, and increment/
decrement the value with the up/down arrow keys or the modify knob. The
change is instantly reflected to the output.
Alternatively, enter a numeric value using the ... numeric keypad. Then press the
ENTER key or the [%] unit key (soft key) to set the input value and reflect it to the
output. If the ENTER key is used, the unit is set to %.
b) Relationship between symmetry and waveform
One waveform cycle changes as follows according to the symmetry setting. A symmetry is
expressed by the amount of the ratios of the first and second rises. Except in the case of 0%
symmetry, the phase zero position is fixed to the 0 amplitude center position.
If it is preferable for the phase zero position to be at the bottom of the waveform, use the
bottom referenced ramp wave of a parameter-variable waveform. ) “1.2.7 Other waveform
group” in the Application Instruction Manual
Symmetry input
field
ENTER
0
9
ENTER
100%80%50%20%0%
Phase
zero position
Phase
zero position
Phase
zero position
Phase
zero position
Phase
zero position
Phase
zero position
Example of 20% and 80% symmetry
for bottom referenced ramp wave

Table of Contents

Related product manuals