EasyManua.ls Logo

Nicoya OpenSPR-XT - Instrument Maintenance and Cleaning

Nicoya OpenSPR-XT
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
73
Figure 6.2 - Examples of bad sugar injections where bubbles were present in valve or flow cell.
6.2 Instrument Maintenance and Cleaning
To prevent clogging or contamination in future experiments it is recommended that deionized water
be pumped through the system after the day’s experiments are completed to prevent salt build up in
the lines. Purges with ethanol can also be performed for additional cleaning.
It is recommended to perform more in depth cleaning every 1-4 weeks depending on instrument
usage. The following cleaning procedure is recommended:
1. Load a clear glass chip (provided) or simply load a used Sensor Chip backwards so the sensor
is not in contact with the Flow Cell.
2. Use DI water as the running buffer, and follow the software prompts (reference spectra dont
matter in this case), ensuring that Automatic Fill of Flow Cell is selected, which will run the
pump at max flow rate. Do not perform bubble prevention IPA injections (skip these steps).
3. In a 96-well plate, prepare 300 µL of 0.5% sodium dodecyl sulphate (SDS), 6 M urea, 1% acetic
acid and 0.2 M sodium bicarbonate solutions into separate wells. Enter the corresponding
sample information and well positions into the Sample Details screen with a pump speed of
25 µL/min for each sample.
4. Prepare the Test Timeline according to the following procedure:
Step
Sample
Timepoint (s) to
proceed to next step
Start
--
50
1
0.5 % SDS
950
2
6 M Urea
1850
3
1% acetic acid
2750
4
0.2 M sodium bicarbonate
3650

Related product manuals