Operation Procedure
5 – 5 – 2 Link Module
The link function enables two systems (NSK VarioSurg 3 and Surgic Pro2) to be controlled by one foot control�
Connecting Link Module to the VarioSurg 3 enables the function to link with this device to be used�
See Link Module manual for the method of use�
1
Turn on the VarioSurg 3�
2
Press and hold the button on Link Module�
3
When the LED turns blue, release the button�
4
Pairing is successful when Surgic Pro2 beeps and
displays “OK”�
Make sure
icon below is displayed and the device
is connected� Also the LED of the Link Module turns
green�
5
Press twice to end setting mode�
CAUTION
If any abnormalities are found with VarioSurg 3 and Link Module, refer to the respective operation manuals�
If the button is not released after the LED turns blue, the LED blinks white and pairing stops� To pair, release the
button, and after the LED lights white, proceed from step 2�
NOTICE
Connect the Link Module to the VarioSurg 3 beforehand�