EN-98 EXPLORER BALANCES
4.18 Library
When an item is processed on a regular basis, the item’s data may be stored in memory for future use. This
memory is referred to as the Balance’s Library.
The following data is stored for the Application used:
Application
mode
Library
function
Characters
of record
name
Type of data to be stored in
library
Comments
Number of
library records
Basic
Weighing
25
Minimum Weight, Sample
Name, Sample ID, Batch ID
Only Admin can edit min-
weight value.
1000 records
without limit to
specific
application
Parts
Counting
25
APW, Sample Size, Over Limit,
Under Limit, Target Sample
PCS, Sample Name, Sample ID,
Batch ID
Sample Size can be set
to 1-10000 pieces.
Percent
Weighing
25
Reference Weight, Reference
Adjust, Sample Name, Sample
ID, Batch ID
N/A
Check
Weighing
25
Over Limit, Under Limit, Nominal
Weight, +Weight Tolerance, -
Weight Tolerance, +Percent
Tolerance, -Percent Tolerance,
Sample Name, Sample ID,
Batch ID
N/A
Dynamic/Ani
mal
Weighing
25
Averaging Time, Sample Name,
Sample ID, Batch ID
Averaging Time can be
set to 1-99 seconds.
Filling
25
Target Weight, Sample Name,
Sample ID, Batch ID
N/A
Totalization
N/A
N/A
N/A
EX can keep totalization
records with up to 999
"Accumulate" operations.
Formulation
only in
"Recipe
Based"
mode
N/A
Item #, Item Name, Item Weight
Number of items in one
record can be 2 to 99.
Differential
Weighing
N/A
N/A
N/A
Up to 99 items are
allowed to store in one
record.
Density
Determinatio
n
25
Water Temperature, Oil Density,
Sinker Volume, Liquid Density,
Sample Name, Sample ID,
Batch ID
N/A
Peak Hold
N/A
N/A
N/A
N/A
Ingredient
Costing
N/A
N/A
N/A
EX can keep last item
settings. Up to 99 items
are allowed in one
formula/recipe.
Pipette
Adjustment
25
Nominal Volume, Inaccuracy,
Imprecision, Water
Temperature, Pressure, Liquid
Density, Pipette Name, Pipette
Number
Number of samples can
be set to 2-30 for one
test.
SQC
25
Nominal Weight, +Tolerance 1, -
Tolerance 1, +Tolerance 2, -
Tolerance 2
Up to 10 batches can be
stored. Number of
samples can be set to 2 -
130 for one batch test.
Fill Weight
Variation
N/A
N/A
N/A
EX can keep last item
settings. Number of
samples can be set to 2-
99 for one test.