EXPLORER BALANCES EN-115
Device Settings Menu: (RS232 shown)
5.10.2 Baud Rate
Set the baud rate (bits per second).
1200 =1200 bps
2400 = 2400 bps
4800 = 4800 bps
9600 = 9600 bps
19200 = 19200 bps
38400 = 38400 bps
5.10.3 Transmission
Set the data bits, stop bit, and parity.
7 EVEN 1 = 7 data bits, even parity
7 ODD 1 = 7 data bits, odd parity
7 NONE 1 = 7 data bits, no parity
8 NONE 1 = 8 data bits, no parity
7 EVEN 2 = 7 data bits, even parity
7 ODD 2 = 7 data bits, odd parity
7 NONE 2 = 7 data bits, no parity
8 NONE 2 = 8 data bits, no parity
5.10.4 Handshake
Set the flow control method.
NONE = no handshaking
XON-XOFF = XON/XOFF handshaking
HARDWARE = hardware handshaking (COM1 menu only)
5.10.5 Alternate Command
Enter this sub-menu to set a different command character for the P (Print), T (Tare) or Z (Zero)
commands.
Note: The selected character can only be used for one command.
Alternate Print Command
Set the alternate command character for Print.
Settings of A to Z are available. The default setting is P, except T&Z.
Alternate Tare Command
Set the alternate command character for Tare.
Settings of A to Z are available. The default setting is T, except P&Z.
Alternate Zero Command
Set the alternate command character for Zero.
Settings of A to Z are available. The default setting is Z, except P&T.
Print Settings Menu:
Note: If LEGAL FOR TRADE is set to ON, the print Output-Stable setting is not reset.
5.10.6 Print Output
Print Stable Weight Only
Set the printing criteria.
OFF = values are printed immediately, regardless of stability.

Table of Contents