EasyManua.ls Logo

Olympus TomoScan FOCUS LT - Instrument Connection

Olympus TomoScan FOCUS LT
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA041-01EN [U8778479], Rev. C, May 2014
System Installation
27
2.3 Instrument Connection
This section explains the procedure for connecting the TomoScan FOCUS LT
instrument. The probe connectors are located on the front panel of the TomoScan
FOCUS LT. The
DC IN connector is also located on the front panel. However, the AC
power entry module connector is located on the back panel.
The instrument must be connected according to the manufacturers instructions in
order to prevent risk of electric shock.
The TomoScan FOCUS LT must always be completely inserted into its housing and
the front panel screws properly tightened in order to ensure proper grounding of the
instrument. Improper grounding of the instrument may produce electrostatic
discharges that could damage the electronic components or cause electric shocks.
To connect the TomoScan FOCUS LT instrument
1. Ensure that the TomoScan FOCUS LT is disconnected from the power source.
To prevent the risk of an electric shock, first disconnect the power cord from the
power outlet, and then disconnect the power cord from the instrument.
2. Install the TomoScan FOCUS LT away from heat sources, leaving a minimum
clearance of 5 cm (2 in.) to allow for heat dissipation.
3. Using an Ethernet cable, connect the TomoScan FOCUS LT
ETHERNET connector to
an Ethernet network or to the network interface card of the control and analysis
computer. The
ETHERNET connector, which is an input/output terminal, can
connect directly to a data-acquisition instrument. The cable supplied with the
instrument is by default a category 5e Ethernet crossover cable.
4. Using the appropriate cables, connect one or more probes to the
PHASED ARRAY
connector and/or to the
P125 to P128 UT connectors (for details, see “Probe
Connection” on page 28).

Table of Contents