DMTA041-01EN [U8778479], Rev. C, May 2014
Specifications
45
5.2.1.1 Specifications of Models FLT16128-P-O and FLT16128-R-O
FLT32128-R-O
K ≥ 1 K = L Pulse-echo N ≤ 31 M ≤ 31
K ≥ 1 K ≠ L Pitch-and-
catch
N ≤ 31 M ≤ 31
FLT3232-P-O
K ≥ 1 K = L Pulse-echo N ≤ 31 M ≤ 31
K ≥ 1K ≠ LPitch-and-
catch
N ≤ 15 M ≤ 15
FLT64:128-P-O
K ≥ 1 K = L Pulse-echo N ≤ 63 M ≤ 63
K ≥ 1 K ≠ L Pitch-and-
catch
N ≤ 31 M ≤ 31 Limit 4
K + (L + M) ≤ 64
(n = 1)
FLT64128-R-O
K ≥ 1 K = L Pulse-echo N ≤ 63 M ≤ 63
K ≥ 1K ≠ LPitch-and-
catch
N ≤ 63 M ≤ 63
FLT6464-P-O
K ≥ 1 K = L Pulse-echo N ≤ 63 M ≤ 63
K ≥ 1 K ≠ L Pitch-and-
catch
N ≤ 31 M ≤ 31
Table 7 Specifications of models FLT16128-P-O and FLT16128-R-O
Category Specification FLT16128-P-O FLT16128-R-O
Pulser Pulse voltage 40 V, 80 V, and 115 V, ±10 % (in high impedance)
Pulse width 30 ns to 500 ns, in steps of 2.5 ns,
precision of 5 ns or ±10 %
Pulse shape Negative square pulse
Fall time < 15 ns
Output impedance < 15 Ω < 45 Ω
Table 6 TomoScan FOCUS LT hardware limitations (continued)
Model Mode Limit 1 Limit 2 Limit 3