DMTA041-01EN [U8778479], Rev. C, May 2014
Chapter 6
66
When the instrument is used in conventional ultrasound mode with the P125, P126,
P127, and P128 connectors, the voltage present on the P connectors can be dangerous
as a shock hazard.
6.6 UT PHASED ARRAY Connector
Description
160-pin, female, Minidock-type connector
Manufacturer; number
Global Components; 30033-160T
Olympus; 21AI0170 [U8904017]
Suggested cable connector; number
Framatome; 89649-002
Olympus; 21AI0153 [U8906010]
Table 17 Pinout for the
P connectors
Connector I/O Signal Description
P125 to P128 Input/Output RF The P connectors are used to
transmit and receive the RF signals
of four conventional ultrasound
probes.