DMTA041-01EN [U8778479], Rev. C, May 2014
System Installation
29
2.5 Automatic Startup Modes
The TomoScan FOCUS LT provides the three following automatic startup modes:
Auto-boot on AC power
For more information, see “Auto-Boot on AC Power” on page 29.
Auto-boot on DC power
For more information, see “Auto-Boot on DC Power” on page 30.
•Auto-UT
For more information, see “Auto-UT Mode” on page 30.
These modes are methods for starting the TomoScan FOCUS LT and defining the
UT
ON button startup state when power is applied to its AC or DC power source inputs.
2.5.1 Auto-Boot on AC Power
Use the auto-boot on AC power mode to remotely turn on a TomoScan FOCUS LT
from an AC power source. When this mode is enabled, you do not need to press the
front-panel on/off switch ( ) to start the TomoScan FOCUS LT. The instrument starts
automatically when AC power is applied to the power entry module on the back
panel. The mode is disabled by default.
To toggle the state of the auto-boot on AC power mode
1. Turn off the TomoScan FOCUS LT instrument using the on/off switch ( ) located
on the front panel.
2. Disconnect the AC power cord from the power entry module on the back panel.
3. Press and hold the on/off switch ( ) on the front panel.
4. Reconnect the AC power cord to the power entry module on the back panel.
5. Release the on/off switch ( ) on the front panel when the
STBY indicator light is
green.
6. To toggle the state of the mode, repeat steps 1 to 4 above.