5.2.2 Specifications of Conventional UT Channels
There are four ultrasound connectors (P125, P126, P127, and P128) to connect
conventional probes to the TomoScan FOCUS LT, which correspond to the last four
channels of the
UT PHASED ARRAY connector.
The specifications for these channels are the same ones as the specifications for the
phased array channels.
When the instrument is used in conventional ultrasound mode with the P125, P126,
P127, and P128 connectors, the voltage present on the P connectors can be dangerous
and create a shock hazard.
The four UT channels are not available on models FLT3232-P-O and FLT6464-P-O.