54
PART 5
SPECIFICATIONS
SENSOR SPECIFICATIONS
Relative Humidity (RH)
Accuracy/Range at 25ºC ±2.75% for 10 to 90%;
±3% for 5 to 10% and 90 to 95%
±4% for 0 to 5% and 95 to 100%
Refer to chart in Appendix H
Hysteresis: ±1% RH
Non-linearity: ±3%
Response Time: 8 seconds, tau 63% (time for reaching 63% of a step
function, valid at 25C and 1m/s airflow)
Repeatability: ±0.1%
Resolution: 0.1%, 12bit
NOTE: Reconditioning of the probe may be necessary if
the probe is stored for a period of time in a harsh
environment (temperatures below 0°C or above 70°C or
exposure to chemical vapors, condensation, etc).
To recondition the probe refer to Appendix H.4
Temperature (T)
Accuracy/Range*:
Wand Probe: ±0.5°C for 5° to 45°C (±1°F for 41° to 113°F);
Up to ±1°C for 0° to 5°C and 45° to 70°C
Up to ±2°F for 32° to 41°F and 113° to 158°F)
Refer to chart in Appendix H
Industrial Probe:
±0.5°C for 5° to 45°C ( ±1°F for 41 to 113°F);
Up to ±1.5°C for -40° to 5°C and 45° to 124°C
Up to ±2.7°F for -40° to 41°F and 113° to 255°F)
Refer to chart in Appendix H
*Note: extended temperature range is for Industrial Probe
only, the iServer’s operating temperature is 0 to 60°C
Response Time: 5 to 30 seconds, tau 63% (response time depends on heat
capacity of and thermal resistance to sensor substrate)
Repeatability: ±0.1°C
Resolution: 0.1°C, 14 bit
PROBE SPECIFICATIONS
Wand Probe: 198 mm long x 19 mm diameter (7.8” x 0.75”)
Cable with DB9 connector: 152 mm long (6”)
Cable operating temperature: 0 to 80°C (32 to 176°F)
Industrial Probe: 137mm or 51mm long x 16mm diameter (5” or 2” x 0.63”)
Cable with DB9 connector: 3m or 0.9m long (10’ or 3’)
Cable operating temperature: -40
to 125°C (-40 to 257°F)
- Non-Condensing: