Technical Data
121
Power factor PF (cos ϕ)
Phase angle
ϕ
Impedance Z
Inductance Lx (equivalent serial circuit)
Quality factor QF
Power P, Q, S (selective)
Range Resolution Typ. accuracy Conditions
0 ... 10%
(capacitive)
5 digits
error < 0.1% of reading
+0.005%
1
1. Reduced accuracy of PF at mains frequency or its harmonics. Mains frequency suppression
available by precisely selecting a mains frequency of *50Hz or *60Hz in the "Hz" column.
f=45...70Hz,
I<8mA,
Vtest = 300V ... 10kV
0 ... 100% 5 digits
error < 0.5% of reading
+0.02%
Vtest = 300V ... 10kV
Range Resolution Typ. accuracy Conditions
-90° ... +90° 4 digits error < 0.01° Vtest = 300V ... 10kV
Range Resolution Typ. accuracy Conditions
1k
Ω ... 1200MΩ 6 digits error < 0.5% of
reading
Vtest = 300V ... 10kV
Range Resolution Typ. accuracy
1H ... 1000kH 6 digits error < 0.3% of reading
Range Resolution Typ. accuracy
0 ... 1000 5 digits error < 0.5% of reading + 0.2%
> 1000 5 digits error < 5% of reading
Range Resolution Typ. accuracy
0 ... 3.6kW 6 digits error < 0.5% of reading + 1mW
0 ... 3.6kvar 6 digits error < 0.5% of reading + 1mvar
0 ... 3.6kVA 5 digits error < 0.5% of reading + 1mVA