40
capacitance and R
p
the losses (Figure 5-1). Losses can also be represented by
serial equivalent circuit diagram with C
s
and R
s
(chapter 12.1). The definition of
the dissipation factor and the vector diagram are shown in Figure 5-2.
Figure 5-1 Simplified circuit diagram of a capacitor
I
CP
U C
P
I
RP
R
P