optris LS – E2006-01-A
36
Factory default settings
The unit has the following presettings at the time of delivery:
Emissivity: 0,950 Lock: Off
Optics: SF Buzzer: On
High alarm: 900°C/ deactivated Laser: On
Low alarm: -35°C/ deactivated Display backlight: On
Temperature unit: °C Display turn: Auto
The Reset function will set the unit back to these default values (exception: optics).
Troubleshooting
Display Problem Action
temperature reading: LLLL object temperature below choose target within
measurement range measuring range
temperature reading: HHHH object temperature above choose target within
measurement range measuring range
battery symbol is on or low batteries check/ replace batteries
flashing
blank display empty batteries check/ replace batteries
immediately
laser does not work low batteries/ see above
laser deactivated activate the laser