155
Pneumatic control
Handpiece Pneumatic Pressure Control
§ How to adjust master block
Turning clockwise decreases the pressure.
Turning counter-clockwise increases the pressure.
Perform adjustment while checking the pressure gauge.
Set the pressure to 4 bar.
Scaler
High Speed Handpiece 1
High Speed Handpiece 2
Low Speed Handpiece
Pressure gauge