90
Wiring Test
v
Light ON/OFF
Light PCB How to check Light PCB
Check AC_IN, AC12V voltage - tester
Perform 7A current test on the fuse
Perform visual check on D2 LED (D2 LED is turned
on/off when the Dr. Table membrane light button is
pressed)
Check the light output power voltage (LAMP): AC12V
should be read when turning the light button ON
LIGHT Test
§ Light PCB Check and diagnosis