13
1.7 Requirements for the controlled product
The minimum recommended weight of the controlled product is in table 1.6
Table 1.6
Marking of the probe type
Minimum weight, kg
Minimum weight, lb.
D
5
11
G
15
33
Е
5
11
UCI
0,3
0.67
If the weight of the controlled product is less than the specified - at measurements it is
necessary to be guided in addition by paragraph 1.10 "Measurements on light and thin-walled
products".
The minimum recommended thickness of the controlled area of the product - is
according to table 1.7.
Table 1.7
Marking of the probe type
Minimum thickness, mm
Minimum thickness, inch
D
25
1
G
70
2.8
Е
25
1
UCI
5
0.2
If the thickness of the controlled product is less than specified - when measuring must
be further guided by paragraph 1.10 "Measurements on light and thin-walled products".
The maximum recommended surface roughness of the product (surface area) on which
measurements are made by table 1.8.
Table 1.8
Marking of the probe type
Surface roughness, no more
D
Rа 3,2
G
Rа 7,2
Е
Rа 3,2
UCI
Rа 1,6

Table of Contents