Then a window will open where you need to enter the number of hardness measures
for calibration (Fig. 2.19), use the buttons
to enter the desired number of measures and press .
Figure 2.19 - Enter the number of hardness measures for calibration
A window will appear (Fig. 2.20), where measurements should be made on calibration
measures one after another. After carrying out several impacts on one of the measures of
hardness it is necessary to press the button and in a window (Fig. 2.21) by means buttons
to enter the specific value of a
measure and to press .Then you need to repeat the same steps on the remaining measures.