Instrument Methods . 63
2. Press 6 to open the Kinetics method.
The Kinetics ā Method Settings 1 screen is displayed.
Enter all numerical values using the keypad numbers or the left and right arrows.
3. Set the Wavelength by using the keypad numbers or left and right arrows, and then
press the down arrow.
The default value is 340 nm.
4. Enter the Delay Time in seconds before measurements are taken and then press the
down arrow.
This can be a maximum of 600 seconds (10 minutes).
5. Enter the time in minutes over which measurements are taken (Duration) and then
press the down arrow.
This can be a maximum of 60 minutes.
6. Enter the Interval time in seconds between measurements using the left and right
arrows, and then press the down arrow.
The options are 1, 2, 5, 10, 15, 20, 30 or 60 seconds.