64 . Lambda XLS/XLS+ User's Guide
7. Press Next .
The Kinetics – Method Settings 2 screen is displayed.
8. Select the measurement Mode using the left and right arrows, and then press the down
arrow.
The options are Delta A (the change in absorbance over the measurement duration, or
selected period), Final A (the absorbance at the end of the measurement duration, or
selected time) and Slope (the rate of change of absorbance over the measurement
duration or selected period).
9. Enter the Units and then press the down arrow.
You can enter a text string up to 8 characters long. To access a list of pre-defined units
on the Units screen, press Options
and then use the left/right arrows (µg/ml,
µg/µl, pmol/µl, mg/dl, mmol/l, µmol/l, g/l, mg/l, µg/l, U/l, %, ppm, ppb, conc or
none). These units can also be edited once OK is pressed.
The Units screen also enables you to select the number of displayed decimal points
(DP). The options are Auto, 0, 1 and 2.
NOTE: The result will always be fixed to 5 significant figures, regardless of how many
decimal points are selected (so 98768.2 is displayed as 98768, even with 1 decimal
point selected).
Press OK
to save your options and return to the Method Settings screen.
You can still edit the units.