EasyManuals Logo

PerkinElmer SPECTRUM TWO N User Manual

PerkinElmer SPECTRUM TWO N
125 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #46 background imageLoading...
Page #46 background image
46 . Spectrum Two N User's Guide
Moving the Spectrometer
WARNING
AVERTISSEMENT
Before moving the spectrometer, switch off the power supply, and
disconnect the power cable.
Avant de déplacer le spectromètre, éteignez l'alimentation et
débranchez le câble d'alimentation.
The spectrometer can be lifted using the shaped handholds on its sides, as shown in
Figure 28.
Figure 28 Lifting the Spectrum Two N
For transportation between a car or laboratory, or between laboratories within your
establishment, we recommend that you use the Spectrum Two Carry Case (part number
L9004199).
If you need to ship your spectrometer, or return it to PerkinElmer following instrument
replacement, you should use the original packaging.
NOTE: Refer to
Appendix 3: Decontamination
and Cleaning on page 120 for more
information.
Although the Spectrum Two N spectrometer is usable within 15 minutes of being switched on
after being switched off overnight, if your instrument has recently been transferred from a
much colder environment it is advisable to run a background spectrum before every sample
for the first couple of hours of use.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer SPECTRUM TWO N and is the answer not in the manual?

PerkinElmer SPECTRUM TWO N Specifications

General IconGeneral
BrandPerkinElmer
ModelSPECTRUM TWO N
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals